8.1 Troubleshooting

114

780/781 pH/Ion Meter, Manual

Problem

Possible cause

Remedy

Measuring signal absent
or very variable

Electrode not connected

Connect electrode

Air in or in front of the dia-
phragm

Remove air

Faulty electrode

Replace electrode

Measured value remains
unstable and does not
fulfill drift criterion

Dirty pH or ISE membrane or
diaphragm

Clean membrane or diaphragm

pH or temperature of the solu-
tion is not stable

Measure under exclusion of air

Thermostat the solution

Unsuitable electrode:

Conductivity too low

Organic solution

Use suitable electrode

Electrode not connected

Connect electrode

Faulty electrode

Replace electrode

Measured value reached
slowly

Dirty pH or ISE membrane or
diaphragm

Clean membrane or diaphragm

Slope too small on cali-
bration

pH or ISE membrane or dia-
phragm is dirty

Clean membrane or diaphragm

Glass membrane dehydrated
after measurement in non-
aqueous solution

Immerse electrode in water between

measurements

Poor buffer/standard solutions

Replace buffers or standards

Electrode "used up"

Replace electrode

Measured value obvious-
ly incorrect

Incorrect calibration

Check / Repeat calibration

Check / Replace buffer/standard so-

lutions

Manual temperature input incor-
rect

Enter correct temperature of the solu-

tion

Dirty pH or ISE membrane or
diaphragm

Clean membrane or diaphragm

Electrolyte or electrode too old

Refill electrolyte or replace electrode

Faulty electrode

Replace electrode