8 Troubleshooting – Messages – Maintenance

780/781 pH/Ion Meter, Manual

115

Problem

Possible cause

Remedy

With standard addition
'auto' the potential step
'delta U' is not reached

Streaming potential of the
measuring or reference elec-
trode causes potential differ-
ence between addition in stirred
and measurement in unstirred
solution

Lower dosing rate

Increase stirring rate

Increase 'delta U'

'auto dos' instead of 'auto'

System test message:
system error #X

Instrument fault

Note error number and inform

Metrohm Service

8.2

Messages

The pH/Ion Meter uses various messages to selectively inform you

about possible errors or operating problems. These are shown in the

bottom line of the display and illustrated by a symbol in front:

Type

Symbol

Example of a message

Error

same buffer

Warning

cal.data out of limits

Information

manual stop

change buffer <ENTER>

calibration OK

Monitoring

cal.interval expired


These messages can normally be removed by pressing the <QUIT>

key. However, some messages will remain until their cause has been

remedied. For example, this applies to a monitoring message or a

warning message during a standard addition run. In the latter case it is

necessary to carry out a new and successful analysis to remove such a

warning message.

Please note that you should not press <QUIT> to acknowledge a

message while a report is being produced as this will cancel the re-

port.

The following table lists all the messages which could appear on the

display of the pH/Ion Meter.