7.1 Calibration and addition/subtraction data

88

780/781 pH/Ion Meter, Manual

reset cal
With this option you can reject all the calibration data of an elec-

trode Id. They are then in the same state as before the first cali-

bration (slope = 100%, pH(0) = 7).

Although the electrode Id is retained, it cannot be selected under

PARAMETER pH/measuring parameters/electr.id

until it

has been calibrated again.

b1, b2...bn
For each calibration buffer the pH value used and the measured poten-

tial are given. The given value for the pH is taken from the stored series

of buffers of the buffer type selected in the calibration parameters (see

Section 9.4). The pH values for temperatures within the given 5 °C steps

are calculated by linear interpolation.
A buffer which has been deleted by

cal.table: delete n

is shown

here by '

deleted

'.

curve

< –> >

If at least 2 calibration buffers have been used then the calibration

curve will be shown as a graph here. The graphics display is accessed

with <

>:

This display is exited with <QUIT>.