6.2 pH measurement (pH mode)

62

780/781 pH/Ion Meter, Manual

6.2.3

Limits pH

PARAMETER pH

:

limits pH

: └

state

In the direct measuring modes the pH/Ion Meter has a limits function

which allows the control of the parameter to be monitored. The control

signals are transmitted via the MSB connection to the optional remote

interface (6.2148.010) (see Section 9.5). Out of limits conditions are al-

so shown on the display and documented in the report.

state:

ON, OFF

u.limit pH

-19.999...14.000...19.999

u.hyst. pH

-19.999...0.020...19.999

l.limit pH

-19.999...0.000...19.999

l.hyst. pH

-19.999...0.020...19.999

The exact way in which the limits function works is described in Sec-

tion 7.4 independent of the mode.

6.2.4

Limits T

PARAMETER pH

:

limits T

: └

state

As well as the primary measuring quantity pH, the temperature can also

be controlled by the limits function. However, you should only activate

the limits function for the temperature when a temperature sensor is

connected, as otherwise an error message will appear.

state:

ON, OFF

u.limit

-999.9...100.0...999.9 °C

u.hyst.

-999.9...0.2...999.9 °C

l.limit

-999.9...0.0...999.9 °C

l.hyst.

-999.9...0.2...999.9 °C

A detailed description of the remote lines is given in Section 9.5. The

exact way in which the limits function works is described in Section 7.4

independent of the mode.