Metrohm 781 pH/Ion Meter User Manual

Page 71

Advertising
background image

6 Methods / Parameters

780/781 pH/Ion Meter, Manual

61

buffer type:

mixed

A further way of using the stored buffer table is to combine a

maximum of five different buffer types to form a new series. You

can make your selection from all the stored buffers described in

Section 9.4.

A general rule is that you can only calibrate in defined temperature

ranges, as otherwise the error message:

buffer not defined

appears. If you use more than two buffers for the calibration then you

can use certain buffers several times in order to give them a better sta-

tistical weight.

low lim.slope

0.1...95.00...999.9 %

up lim.slope

0.0...103.0...999.9 %

low lim.pH(0

-99.999...6.40...99.999

up lim.pH(0)

-99.999...8.00...99.999

)The most important calibration data such as slope and pH(0) must lie

within the limits defined here in order for them to be automatically

adopted. The given default values for the slope correspond to the quali-

ty criteria of a "passing electrode" in the electrode test (see Section 8.6).
If one of the values of the calibration data lies outside these limits then

at the end of the calibration the following message will appear:

cal.

data out of limits

; you can then accept or reject the results.

offset Uoff state: OFF, ON

offset Uoff

-2200.0...0.0...2200.0 mV

The automatic buffer recognition of the pH/Ion Meter assumes an

offset potential of 0 ± 30 mV for the electrode used. Ag/AgCl ref-

erence electrodes normally meet this requirement. If other refer-

ence systems are used, e.g. a calomel electrode, then the offset

potential may be outside this range. This means that for correct

buffer recognition this variation must be compensated by entering

the specific offset potential Uoff. For the pH measurement itself

this parameter is unimportant.

Advertising
This manual is related to the following products: