Metrohm 781 pH/Ion Meter User Manual

Page 86

Advertising
background image

6.5 Direct ion measurement (781 only: Conc mode)

76

780/781 pH/Ion Meter, Manual

drift:

0.1...0.5...9.9 mV/min

Just like the ion measurement, the calibration is also drift-controlled.

However, this drift-control cannot be switched off.

report:

OFF, short, full

After the calibration you can automatically output the result in the form

of report. You can choose between the '

short

' version containing all the

relevant data and the '

full

' version, which also shows the measuring

curves as a graph (see Section 7.2.4).
Such a report can also be printed out later together with the calibration

for each stored electrode Id.

calibration interval

OFF, 1...999 h

The pH/Ion Meter can automatically remind you about any recalibration

of the pH electrode which is due. This means that if a calibration is car-

ried out with a preset calibration interval then, when this interval has

elapsed, a message will appear on the display:

cal.interval expired

This message is retained and will be printed out on every measuring

point report until a new calibration has been carried out.

no.of standards 1, 2...19
Up to 19 standards can be used for the calibration of the ion selective

electrode. If only a single standard is used for the calibration then the

program will use the existing value for the slope. This will also be cor-

rected accordingly if the calibration temperature changes. This adop-

tion of the existing slope also applies when a calibration with several

standards is canceled with <MODE> or <QUIT> after the first stand-

ard has been measured and then confirmed with <ENTER>.
However, please note that for the calibration of ion-selective electrodes

at least two standards should be used. The possibility of using a single

standard described above should only represent an absolute excep-

tion.
If three or more standards are used then the variance will be included in

the calibration data. More detailed information about the evaluation is

given in Section 9.2.

addition:

manual, auto

For the manual addition of the standard you should enter in the next

calibration parameter lines the concentrations of the standards used

(Conc.1, Conc.2, ...). This data can still be modified during the cali-

bration sequence.
With automatic standard addition the pH/Ion Meter carries out the cal-

culation of the standard concentrations, the preparation of the standard

solutions and their measurement automatically. A requirement is that a

Metrohm Dosimat Plus is connected (connection and Dosimat Plus