7.2 Reports

102

780/781 pH/Ion Meter, Manual

7.2.5

Result report (781 pH/Ion Meter only)

In the Conc mode for standard or sample addition the results can be

produced as a result report. Under

PARAMETER Conc/Standard ad-

dition/Report

you can define whether this result is to be produced

automatically after the end of the measurement. Just as for the calibra-

tion report, you can choose between a short version containing all the

important data and a long version which additionally contains a graph-

ical presentation of the evaluation.

'ca

781 pH/Ion Meter

01104 5.781.0020

date 2004-06-03 09:38:25

user

C. Weber

Conc F-StdAdd run number

2

electr.id

F-electr.011

addition/subtraction method

meas.type: std.add

temperature 20.1 °C

temp.sensor Pt1000

conc.std 1000.0 ppm

V total 20.0 mL

initial voltage 180.0 mV

factor 25.0

smpl.size 10.0 mL

date 2004-06-03 09:33

dV/ml U/mV dU/mV

std.incr. 1 0.020 115.5 -64.5

std.incr. 2 0.026 95.5 -20.0

std.incr. 3 0.056 76.0 -19.5

std.incr. 4 0.123 55.6 -20.4

variance 0.080

slope -59.0 mV

E(0) 117.7 mV

c F(-1) 4.40 ppm

==============

REPORT

9

+

ADD.DATA

4