7.2 Reports

106

780/781 pH/Ion Meter, Manual

7.2.9

User method report

Method and calibration data occupy memory space in the pH/Ion Me-

ter. The user method report gives an overview of the memory occupan-

cy:

'um

781 pH/Ion Meter

01104 5.781.0020

date 2004-06-03 13:58:49

user memory

>methods

pH pH-Demo 82

Conc F(-1) F-Determ 64

>caldata

pH pH 1.1 130

Conc F(-1) F-Sens 146

remaining bytes 4578

--------------

The mode, method name and the occupied memory is given in bytes

for each of the methods. For calibration data the mode, electrode Id

and memory occupancy are given.
A total of 5000 bytes is available for this data. This means that you can

store a maximum of approx. 50 methods or calibration data. This num-

ber can vary depending on the size of the individual data records.
The size of the measured value memory (see Section 7.3) is independ-

ent of this memory administration.