6 Methods / Parameters

780/781 pH/Ion Meter, Manual

77

configuration see Section 2.3.3, pH/Ion Meter configuration see Sec-

tion 5.6).

min.conc

1.0E-30...0.1...1.0E+30

max.conc.

1.0E-30...1.0...1.0E+30

The standard concentrations to be calculated lie within the range of

these minimum and maximum concentrations. Depending on the re-

quired number of standards (see above), the pH/Ion Meter will calculate

the other concentrations. The expected resulting potential difference

between all the standards remains constant.

V init

0.001...100.0...999.9 mL

The initial volume of a presentation (e.g. buffer, TISAB, etc.) is entered

here.

no.of Exchange Units. 1...5

It is normally a good idea to calibrate ion-selective electrodes over sev-

eral concentration decades. It is possible that a single Dosimat ex-

change unit may not be sufficient for this. In this case you can prepare

several exchange units with the corresponding different concentrations

of standard solutions:

conc.1...n

1.0E-30...100.0...1.0E+30

V Exchange Unit 1...n: 1, 5, 10, 20, 50 mL

For each exchange unit you must enter the concentration of the solution

it contains and the volume of the dosing cylinder. Arrange the exchange

units with increasing concentration of the standard solutions.
Please make sure that all exchange units have been prepared before

the start of a calibration sequence (dosing cylinders freshly filled, air

bubbles removed, etc.). The request for changing the exchange unit is

given on the display. This message describes the required standard

concentration and volume increment at the Dosimat Plus.

fill dosimat, dV/dt max.
dV = 1 ul, c = 2000.0 ppm

After mounting and filling the prepared exchange unit, continue with

<ENTER>.

6.5.6

Limits Conc

(applies only to measurement type: direct)

PARAMETER Conc

:

limits Conc

: └

state

In the direct measuring modes the pH/Ion Meter has a limits function

which allows the control of the parameter to be monitored. The control

signals are transmitted via the MSB connection to the optional remote