6.6 Standard and sample addition (781 only: Conc mode)

82

780/781 pH/Ion Meter, Manual

TIP !

It is possible to alter this parameter during an addition run. For

example, before confirming the volume increment you can add

the solution during the run until a certain potential difference is

shown on the display. Then confirm the volume to be added; the

corresponding potential will be measured after each addition.

This can be a great help, for example when checking the suita-

bility of a standard for this method.

stop V

0...99.99...9999.9 mL

For the manual addition no check will be carried out whether the

stop volume is exceeded or not in contrary to

auto dos

and

auto

(see below).

auto dos and auto
If a Dosimat Plus is used then the standard or sample addition

can be carried out automatically (connection and Dosimat Plus

configuration see Section 2.3.3, pH/Ion Meter configuration see

Section 5.6). You can chose between two methods: under auto

dos you can enter the individual volume increments for the add-

ed solution; under auto these additions are made automatically

so that a constant potential difference always results.

The common parameters are:

dos.rate

fast, medium, slow

no.of additions 1, 2, 3,...19
V Exchange Unit 1, 5, 10, 20, 50 mL
stop V

0...99.99...9999.9 mL

The dosing rate can be set in three steps via

dos.rate

.

In '

auto dos

' the pH/Ion Meter checks at the start of the auto-

matic standard and sample addition whether the sum of the vol-

ume increments exceeds the set stop volume

Stop V

. If this is

the case then the following message will appear:

V add too large

Under '

auto

' this check is carried out during the automatic ad-

dition and, if necessary, will cause the run to be stopped with the

message:

check working cond.

Depending on the automatic addition method the following pa-

rameters must also be defined: