5 setup, 1 locking, Setup – Metrohm 781 pH/Ion Meter User Manual

Page 121: Locking

Advertising
background image

7 Various functions

780/781 pH/Ion Meter, Manual

111

7.5

Setup

Certain settings of the pH/Ion Meter can be found in the Setup menu.

This menu cannot be accessed from the normal instrument display, as

alterations to these settings have a great influence on the functions of

the pH/Ion Meter.
The Setup menu is called up immediately after the instrument has been

switched on by pressing down and holding down the <CONFIG> key.

setup

lock

input assign

graphics

instr.number

Be very careful when making alterations in this menu! Alterations have

a very large influence on the functions of the instrument. Accidental

alterations can be rejected by carrying out a deliberate memory initial-

ization (see Section 8.5).

7.5.1

Locking

setup

lock

: ├

all keys

<CONFIG>

<PARAM>

<CAL>

recall method

store method

delete method

<CAL.DATA>

<MODE>

<EL.TEST>

You can disable certain keys and functions, e.g. to prevent unwanted

alterations to the instrument settings or method parameters.
You can activate such a disablement by setting the corresponding pa-

rameter to '

ON

'. This does not affect access to the Setup menu in any

way.
Please note that configuration settings (i.e.

CONFIG/report/id1

and

id2)

and parameters (i.e. manual calibration concentrations or stand-

ard additions), which are edited during a measurement or calibration

procedure, are modified that way even if <CONFIG> or <PARAM>

are locked. The advantage is that the requested settings and parame-

ters only have to be confirmed during repeated procedures.
Please note that all disablements will be invalidated if a RAM-Init is car-

ried out (see Section 8.5).

Advertising
This manual is related to the following products: