8 Troubleshooting – Messages – Maintenance

780/781 pH/Ion Meter, Manual

117

Message

Situation

Cause

Remedy

limit error

Direct measure-
ment, all modes

Set limit for primary or sec-
ondary measured value
exceeded

see

Section 7.4

cal. interval
expired

pH and Conc
modes

Set calibration interval has
expired

Recalibrate electrode (see

Section 6.2.2

and

6.5.5

)

cal.data out of
limits

pH calibration

Determined calibration
data are outside limits de-
fined as calibration param-
eter

Accept calibration data with

<ENTER>

or reject them

with

<MODE>

; check

electrode, buffer or limits
under

6.2.2

Conc calibration

Determined slope is

outside limits of 80 –
105 % of theoretical
slope (59.16 mV per ion
charge [25°C])

Sign incorrect for set

type of ion.

Accept calibration data

with <ENTER> or reject
them with <MODE>



Check electrode, stand-

ards under 6.5.5 or check
ion type under 6.5.2

calibration OK pH calibration,

Conc calibration

Calibration has been con-
cluded successfully

Exit display with

<QUIT>

;

this happens automatically
after 30 s; calibration data
will be accepted

manual stop

Various measur-
ing runs

Measurement has been
stopped manually

measuring
range exceeded

Various measur-
ing runs

Electrode not immersed

in the solution

Electrode not connect-

ed correctly

Remedy fault or change
mode

measured value
memory full

Saving measured
values in a direct
measurement

Attempt made to store a
further measured value
when 100 have already
been stored

Delete measured values
(see

Section 7.3

)

plot data
overflow

Report output

Data output is faster than
printer processing speed

Press

<QUIT>

and

Increase time interval be-

tween measurements

Increase time scale

Reduce graph width

Only output primary

measured value.

buffer not
defined

pH calibration

pH of a buffers has not
been defined under Buff-
er type: own
for the
given temperature

Enter missing value in tem-
perature table (see

Sec-

tion 6.2.2

)