5.2 Printing out measured values

44

780/781 pH/Ion Meter, Manual

5.2

Printing out measured values

MEAS/

PRINT

CONFIG

:

print meas.value

: └

print crit

immediate

date & time

report header

calibration report

time

date & time

report header

calibration report

time interval

stop time

drift

date & time

report header

calibration report

plot

time interval

time scale

time scale label

stop time

OFF

In the configuration menu

print meas.values

the output form of

measured values, triggered by <MEAS/PRINT>, is defined. This is only

possible for the direct measuring modes pH, T, U and Conc(direct).

print crit:

immediate, time, drift, plot, OFF

Depending upon the print criterion a measured value report with be

transmitted via the RS232 interface to a printer or PC either immediate-

ly, at defined time intervals or only when a drift condition has been ful-

filled. If a stirring sequence is intended before the measurement (

PA-

RAMETERS/measuring parameters/stirrer: control,

see Sec-

tion 6), then these print criteria will only be effective when the stirring

sequence has been concluded.
A further way of documenting the measured values is to output a

graphical plot of the measure value signals.

print crit.: immediate
With this setting a measured value report will be produced when

<MEAS/PRINT> is activated from the normal instrument display. The

following options can be used:

date & time:

ON, OFF

Here you can choose whether the date and time at which the

measured value was recorded are to be included with each indi-

vidual measured value. This can take place irrespective of wheth-

er a report header is used.

report header: once, always, OFF