2 functions of the individual remote lines, Functions of the individual remote lines – Metrohm 781 pH/Ion Meter User Manual
Page 167
9 Annex
780/781 pH/Ion Meter, Manual
157
9.5.2
Functions of the individual remote lines
Name
Pin
Function
Input 0
21
Print/Start
Outputs a measuring points or results report with
stepping impulse (Output 3)
t
Puls
> 100 ms
I
nac
tiv
e
M
od
e pH
M
ode
T
M
ode
U
pH
C
al
E
l. T
es
t
M
od
e C
on
c
Co
nc
Ca
l
<EN
TER
>
Input 1
9
= active (low)
t
Puls
> 100
ms
Input 2
22
Input 3
10
Input 4
23
Input 7
12
Ready
"Ready" signal from 865 or 876 Dosimat Plus
(Conc mode: automatic calibration or standard /
sample addition)
Output 0
5
Ready
Instrument is in measuring mode and all runs are
finished
Output 1
18
Prim. meas. value: if above upper limit: active (low)
Output 2
4
Prim. meas. value: if below lower limit: active (low)
Output 3
17
Stepping impulse after successful measurement, t
Puls
> 80 ms
Output 4
3
Stirrer control
Output 5
16
Output 6
1
Sec. meas. value: if above upper limit: active (low)
(781: with autom. addition: FILL for 865/876, t
Puls
>10 ms)
Output 7
2
Sec. meas. value: if below lower limit:
active (low)
(781: with autom. addition: impulse for Dosimat Plus, t
Puls
= 1.25
ms)
+5 V
15
I ≤ 40 mA, R
i
≈ 12 Ω
0 V: active (low), 5 V: inactive
0 V
14/25