2 functions of the individual remote lines, Functions of the individual remote lines – Metrohm 781 pH/Ion Meter User Manual

Page 167

Advertising
background image

9 Annex

780/781 pH/Ion Meter, Manual

157

9.5.2

Functions of the individual remote lines

Name

Pin

Function

Input 0

21

Print/Start

Outputs a measuring points or results report with

stepping impulse (Output 3)

t

Puls

> 100 ms

I

nac

tiv

e

M

od

e pH

M

ode

T

M

ode

U

pH

C

al

E

l. T

es

t

M

od

e C

on

c

Co

nc

Ca

l

<EN

TER

>

Input 1

9

 = active (low)

t

Puls

> 100

ms

Input 2

22

Input 3

10

Input 4

23

 

Input 7

12

Ready

"Ready" signal from 865 or 876 Dosimat Plus

(Conc mode: automatic calibration or standard /

sample addition)

Output 0

5

Ready

Instrument is in measuring mode and all runs are

finished

Output 1

18

Prim. meas. value: if above upper limit: active (low)

Output 2

4

Prim. meas. value: if below lower limit: active (low)

Output 3

17

Stepping impulse after successful measurement, t

Puls

> 80 ms

Output 4

3

Stirrer control

Output 5

16

Error (see Section 8.2)

Output 6

1

Sec. meas. value: if above upper limit: active (low)
(781: with autom. addition: FILL for 865/876, t

Puls

>10 ms)

Output 7

2

Sec. meas. value: if below lower limit:

active (low)

(781: with autom. addition: impulse for Dosimat Plus, t

Puls

= 1.25

ms)

+5 V

15

I ≤ 40 mA, R

i

≈ 12 Ω

0 V: active (low), 5 V: inactive

0 V

14/25