Metrohm 781 pH/Ion Meter User Manual

Page 37

Advertising
background image

3 Short operating tutorial

780/781 pH/Ion Meter, Manual

27

'fc

781 pH/Ion Meter

01104 5.781.0010

date 2004-05-13 09:20:50

Conc ******** run number

1

electr.Id

conc.calibration

ion type F(-1)

temperature 22.7 °C

temp.sensor Pt1000

c.dat. 2004-05-12 09:32

conc/ppm U/mV dconc/%

std. 1 5.00E-01 139.9 0.2

std. 2 2.50E+00 97.0 0.1

std. 3 5.00E+00 78.8 -0.6

std. 4 2.50E+01 37.1 -0.2

std. 5 5.00E+01 19.2 0.4

variance 0.014

slope -59.9 mV

E(0) 120.8 mV

c(blank) 0.00E+00 ppm

==============

Advertising
This manual is related to the following products: