background image

7 Various functions

780/781 pH/Ion Meter, Manual

101

7.2.4

Calibration report

In the pH and Conc(direct) modes a calibration report can be produced

after a calibration has been carried out. Under

PARAMETER/calibra-

tion parameters/report

you can define whether this report is to be

produced automatically after the measurement. You can also choose

between a short version containing all the important data and a long

version which additionally contains a plot of the calibration function.

'fp

781 pH/Ion Meter

01104 5.781.0020

date 2004-06-02 17:10:16

user

C. Weber

pH pH-1 run number

9

electr.id

pH electr.01

pH calibration

temperature 20.1 °C

temp.sensor Pt1000

c.dat. 2004-06-02 17:06

buffer type Metrohm

pH U/mV dpH*)

buffer 1 7.019 0.1 -0.015

buffer 2 3.991 176.7 0.006

buffer 3 9.037 -115.4 0.009

variance 1.192

slope 99.44 %

pH(0) 7.035

==============

*) dpH: Nominal pH value of the buffer – pH value which, at the

mesured voltage, results from the regression analysis

REPORT

9

+

CAL.DATA

7