4 validation, 1 electronic tests, Validation – Metrohm 781 pH/Ion Meter User Manual

Page 132: Electronic tests

Advertising
background image

8.4 Validation

122

780/781 pH/Ion Meter, Manual

8.4

Validation

The checking and maintenance of the pH/Ion Meter can be carried out

in 3 steps. This list can be extended by the measures described in Sec-

tion 8.3.

1.

Checking the electronic components when the instrument is

switched on.

2.

Validation of the whole analytical system by wet chemistry.

3.

pH/Ion Meter maintenance by Metrohm Service.

8.4.1

Electronic tests

After the pH/Ion Meter has been switched on an electronic system test

is carried out automatically. The following display is seen:

The tests are documented in the system test report; this can be printed

out when the instrument is switched on: (see Section 5.4):

'di

781 pH/Ion Meter

01104 5.781.0020

date 2004-05-21 15:32:00

RAM test OK

EPROM test OK

real time clock OK

COMPort OK

A/D converter OK

LCD display OK

==============

If one of these tests is not concluded with '

OK

' then please contact

Metrohm Service.
With a

real time clock

error you can try resetting the date and time.

If the test is then '

OK

', you should check whether the stored methods

and instrument configuration are still unchanged.