7 electrode test, Electrode test – Metrohm 781 pH/Ion Meter User Manual

Page 74

Advertising
background image

6.2 pH measurement (pH mode)

64

780/781 pH/Ion Meter, Manual

6.2.7

Electrode test

PARAMETER pH

:

electrode test

electrode type

temperature

report

The pH/Ion Meter can carry out a pH electrode test, which permits the

quality of the electrode used to be evaluated. A detailed description of

the electrode test is given in Section 8.6.

electrode type: standard, gel, non-aq., own
Various types of pH electrodes have different evaluation criteria. For the

three most commonly used Metrohm pH electrode types the tolerances

are stored in the pH/Ion Meter: standard electrodes, gel electrodes and

non-aqueous electrodes.

electrode type: own
If you wish to define your own tolerances then select '

own

'. You

can now enter the limits for the quality features '

excellent

electrode

', '

good electrode

' and '

passing electrode

'.

An explanation of the individual limits is given in Section 8.6.3.

Regardless of the electrode evaluation, if '

own

' is selected as the

electrode type then you can also edit the limits for the offset po-

tential Uoff. This allows the electrode test to be carried out with

pH electrodes which, even with an Ag/AgCl reference electrode,

show a clearly shifted zero potential (e.g. antimony electrode).

The use of other reference electrodes can also result in changed

zero potentials.

temperature

0.0...25.0...99.9 °C

If no temperature sensor is connected then enter the temperature at

which the buffer will be measured here.

report:

OFF, short, full, line

After the electrode test you can automatically output the result in the

form of report. You can choose between the 'short' version containing

all the relevant data and the 'full' version, which also shows the measur-

ing curves as a graph. If you are only interested in the evaluation of the

electrode itself then select '

line

'. Only the result will be printed out, e.g.

'

excellent electrode

'. If no printer is connected then please select

'

OFF

'.

The report of the last pH electrode test can also be printed out later with

<REPORT> (see Section 4.3).