5 Configuration

780/781 pH/Ion Meter, Manual

47

5.3

Storing measured values

STORE

5

CONFIG

:

store meas. value

: └

store crit

immediate, drift, OFF

time

time interval

stop time

In the direct measuring modes pH, T, U and Conc(direct) you can use

<STORE> to store up to 100 measured values with additional infor-

mation in the instrument memory (see Section 7.3). The functions of the

settings to be defined under

CONFIG/store meas. value

correspond

to those made under

CONFIG/print meas. value

:

store crit:

immediate, time, drift, OFF

This storage criterion also only becomes effective when an optional stir-

ring sequence has been concluded.
In contrast to the report output, the combination of the information

stored together with each measured value is fixed.

store crit:

immediate

The current measured value is stored with <STORE>.

store crit:

time

time interval:

0.1...4.0...99999.0 s

stop time:

OFF, 1...999999 s

You can automatically record and save the measured values at fixed

time intervals until the stop time has been reached.

store crit:

drift

A measured value can also be stored when the drift condition defined

under

PARAMETERS/measuring parameters/drift

has been fulfilled.

After 100 measured values have been stored, and each time a further

storage attempt is made, the pH/Ion Meter will produce a correspond-

ing warning:

measured value memory full

You now have the possibility of using <RECALL> to view the stored

measured values, of deleting individual measured values or of clearing

the whole measured value memory (see Section 7.3). It is also possible

to output all measured values as a report via the RS232 interface (see

Section 7.2.8).