Connecting a computer

The report output via the RS232 interface can be made to either a print-

er or a PC. With the program Metrodata Vesuv 3.0 for Windows it is

possible to implement automatic data acquisition and evaluation.
This is done by connecting RS232 interface 10 of the pH/Ion Meter us-

ing the optionally available 6.2134.040 Cable (9-pin/9-pin) with a free

serial COM port on the PC. If only a 25-pin COM port is available then

the optionally available 6.2125.110 Cable (9-pin/25-pin) or a commer-

cially available adapter must be used.
Settings
The RS232 interface must be configured with identical settings on both

the connected instruments (PC and pH/Ion Meter). You should select

the standard values for the pH/Ion Meter configuration under

CON-

FIG/RS232 settings

(see Section 5.7):

Baud Rate: 9600
Data Bit: 8
Stop Bit: 1
Parity: none
Handshake: HWs

Please note that, before starting the program 'Vesuv Datalogger', the

pH/Ion Meter must have been configured correctly, connected up and

switched on. When the Vesuv program is started it automatically

checks the configuration of the connected instrument and makes the

corresponding settings for the report output so that data acquisition

functions properly.


As well as the data acquisition function described above, the 780/781

pH/Ion Meter also has numerous remote control possibilities that allow

the complete control of the instrument via the RS232 interface. Howev-

er, as the utilization of this option only makes sense in very special situ-

ations, you can obtain detailed information about the remote control

language of the pH/Ion Meter from your local Metrohm agency in a

separate document (780/781 pH/Ion Meter Operation via RS232;

8.781.1113).