3 measuring parameters, Measuring parameters – Metrohm 781 pH/Ion Meter User Manual

Page 82

Advertising
background image

6.5 Direct ion measurement (781 only: Conc mode)

72

780/781 pH/Ion Meter, Manual

conc.unit:

mol/L, %, ppm, g/L, mg/L,

µg/L, mEq/L, own

The analyte concentration unit is defined here. It is used in both the cal-

ibration and the measurement itself.
Under '

own

' you can also define your own unit with 5 ASCII characters:

unit

5 ASCII characters

Please note that this unit does not form part of the measured value itself

but is only a supplement. This means that it cannot be used for calcula-

tion, for example to compensate for changes in order of magnitude. For

this reason the concentration information in the following method pa-

rameters is given without mentioned the unit.

6.5.3

Measuring parameters

PARAMETER Conc

:

measuring parameters

: ├

electr.id

drift

temperature*

method

delta measurement

stirrer

(* only appears if no temperature sensor is connected)
The measuring parameters describe the settings which are of direct

importance for the measurement.

electr.id:

12 ASCII characters

You can assign any name you want to the connected ion-selective elec-

trode. This electrode Id is then contained in the report header for meas-

ured values. Each time that the electrode is calibrated the resulting cal-

ibration data will be assigned to this entered electrode Id.
In this way you can enter several electrode Ids for each type of ion. This

allows you to switch between various electrodes and to select their as-

sociated calibration data. However, a requirement is that each new

electrode Id is linked with a new calibration. Then you can use <SE-

LECT> under the measuring parameter

electr.id

to select one of the

stored electrode Ids which has been calibrated with the current type of

ion.
In order to delete an electrode Id and the calibration data linked with it

this must be reset under <CAL.DATA> (

Reset cal.

, see Sec-

tion 7.1.2).

drift:

0.1...1.0...999.9 mV/min, OFF

A measuring signal which changes as a function of time is regarded as

being sufficiently constant when its drift drops below the value defined

here. This can be used as criterion for the pH/Ion Meter if a displayed