6.3 Temperature measurement (T mode)

66

780/781 pH/Ion Meter, Manual

This reference value is subtracted from the measured value itself. The

modified measured value is shown together with this value and is taken

into consideration, for example, in checking the limit values (see

PA-

RAMETER T/limits T

). The reference value is not included in the re-

port. However, should you still want to document it then you should al-

so produce a parameter report (see Section 7.2.7).

stirrer:

ON, OFF, control

You can choose whether the pH measurement is to be made while stir-

ring is being carried out or not. A requirement is that a Metrohm stirrer

is connected with a suitable cable (see Section 2.3). If an 8xx series stir-

rer has been connected and the settings made under

CON-

FIG/peripheral units

then the stirring rate can also be selected:

stirring rate

1...5...15

If you want your measurement to be made without stirring but still want

to stir the solution before the measurement then you should select the

control

setting. In this way you can trigger a stirring sequence before

the start of each measurement with <MEAS/PRINT>:

prestir pause

0...99999 s

stir time

0...99999 s

poststir pause

0...99999 s

The prestir and poststir pauses represent waiting periods in which stir-

ring does not take place. The measurement itself only starts when the

complete stirring sequence has ended. Please note that this parameter

is binding for the report output and for storing a measured value. There-

fore when you start a measurement with <MEAS/PRINT> the stirrer will

be switched on and off accordingly. This is why you should preferably

carry out drift-controlled measurements (see Section 5.2 and Sec-

tion 5.3).

6.3.2

Limits T

PARAMETER T

:

limits T

: └

state

In the direct measuring modes the pH/Ion Meter has a limits function

which allows the control of the parameter to be monitored. The control

signals are transmitted via the MSB connection to the optional remote

interface (6.2148.010) (see Section 9.5). Out of limits conditions are al-

so shown on the display and documented in the report.

state:

ON, OFF

u.limit

-999.9...100.0...999.9 °C

u.hyst.

-999.9...0.2...999.9 °C

l.limit

-999.9...0.0...999.9 °C

l.hyst.

-999.9...0.2...999.9 °C