2 conc calibration data (781 ph/ion meter only), Conc calibration data (781 ph/ion meter only) – Metrohm 781 pH/Ion Meter User Manual

Page 99

Advertising
background image

7 Various functions

780/781 pH/Ion Meter, Manual

89

7.1.2

Conc calibration data (781 pH/Ion Meter only)

CAL.DATA

7

CAL.DATA

ion type

electr.Id

method

slope

E(0)

c(blank)

temperature

temp.sensor

c.dat.

cal.interval

variance

no.of standards

cal.table

st1

st2

:

stn

curve

ion type
The analyte ion which has been defined during the calibration under

PARAMETER Conc/ion parameters

is shown here. As with the follow-

ing electrode Id, the current calibration data changes as the ion param-

eter changes.

electr.id
If an electrode Id has been entered or selected before the calibration

under

PARAMETERS Conc/measuring parameters/electr.id

then it

will be shown here. Electrode Ids can be defined for a particular type of

ion. The calibration data will apply to this electrode only and will change

when a new electrode Id is set under the above measuring parameters

(see Section 6.5.5).

method
The calibration parameters set under

PARAMETER Conc/calibration

parameters

form part of the method parameters. The appropriate

method name will be shown here when the calibration data have been

determined using this method and the resulting quantities

slope

,

E(0)

and

c(blank)

have not been altered since. If no method name has

been assigned before the calibration (see Section 6.1), then '

********

'

will appear here.

slope
The slope of the calibration function obtained during the calibration will

be shown here in mV. With no calibration this will be set to the theoreti-

cal value of +59.2 mV or -59.2 mV per ion charge at 25 °C.
This value can be altered manually for test purposes. However, after-

wards the information about the method will be missing from the list