6.5 Direct ion measurement (781 only: Conc mode)

74

780/781 pH/Ion Meter, Manual

rer has been connected and the settings made under

CON-

FIG/peripheral units

then the stirring rate can also be selected:

stirring rate

1...5...15

If you want your measurement to be made without stirring but still want

to stir the solution before the measurement then you should select the

control

setting. In this way you can trigger a stirring sequence before

the start of each measurement with <MEAS/PRINT>:

prestir pause

0...99999 s

stir time

0...99999 s

poststir pause

0...99999 s

The prestir and poststir pauses represent waiting periods in which stir-

ring does not take place. The measurement itself only starts when the

complete stirring sequence has ended.
Please note that this parameter is binding for the report output and for

storing a measured value. Therefore when you start a measurement

with <MEAS /PRINT> or an addition sequence with <ENTER> the stir-

rer will be switched on and off accordingly. This is why you should pref-

erably carry out drift-controlled measurements for the direct determina-

tion of the concentration (see Section 5.2 and Section 5.3).

6.5.4

Calculation parameters

PARAMETER Conc

:

calculation parameters

: ├

smpl.size

V total

factor

smpl.size unit

The calculation parameters can be used for directly including sample

preparation data for the automatic calculation of the analytical result. In

this way the measured value shown will refer to the concentration of the

ion in the original sample and not in the sample solution to be meas-

ured.

smpl.size

0.0001...99999.9 (mL or g), OFF

V total

0.001...100.0...9999.9 mL

factor

-1.00E+30...1.0...1.00E+30

smpl.size unit: mL, g

smpl.size

describes the volume or weight of the sample under inves-

tigation, depending on the selected

smpl.size unit

. If you do not

want to use this parameter then set it to '

OFF

'.

V total

represents the volume of the sample solution at the start of

the measurement, e.g. after the addition of a buffer (TISAB, etc.).
The calculation of the measured value is then carried out according to

the following equation: