6 peripheral units, Peripheral units – Metrohm 781 pH/Ion Meter User Manual

Page 62

Advertising
background image

5.6 Peripheral units

52

780/781 pH/Ion Meter, Manual

system test report: ON, OFF
When it is switched on the 780/781 pH/Ion Meter automatically carries

out a system diagnosis. The result can be outputted as a system test

report (see Section 8.4.1).

5.6

Peripheral units

CONFIG

:

peripheral units

: ├

character set

Dosimat

stirrer

keyboard

barcode

Various peripheral units can be connected to the 780/781 pH/Ion Meter.

The corresponding configurations must be carried out here.

character set: IBM, Epson, Seiko, Citizen, HP
The type of printer chosen for the report printout is selected here. If you

have connected a PC to the RS 232 interface for data reception then

set this parameter to '

IBM

'.

Dosimat:

665, 725, 765, 776

The 781 pH/Ion Meter can be used to control a Metrohm Dosimat Plus,

e.g. for the automated production of standard solutions for the calibra-

tion of ion-sensitive electrodes. Analyses involving the addition of

standard and sample solutions can also be automated in this way. In

order to control an 865 or 876 Dosimat Plus, make sure that the

Dosimat 665 is selected. Please note the required settings at the

Dosimat Plus (see Ch. 2.3.3).

stirrer:

8xx, 7xx

Metrohm stirrers of the series 8xx and 7xx can be connected to the

pH/Ion Meter. In order to connect 7xx series stirrers the use of a remote

box (6.2148.010) at the MSB connection of the pH/Ion Meter is neces-

sary (see Section 9.5). For example, both types of stirrer can be

switched on and off from the pH/Ion Meter with <STIRRER>. With 8xx

series stirrers the stirring rate can also be controlled from the pH/Ion

Meter (see

PARAMETERS/measuring parameters/Stirrer

). If you

are using a stirrer of another manufacturer, then you must select

'7xx'

.

keyboard:

US, deutsch, français, español,

schweiz.

If texts have to be entered frequently it may be a good idea to connect

a PC keyboard. The language-specific keyboard layout is then set here.

Older PC keyboards may possibly have a too-high power consumption,

which could result in the cancellation of the system test when the in-