8 Troubleshooting – Messages – Maintenance

780/781 pH/Ion Meter, Manual

119

Message

Situation

Cause

Remedy

RS error 45

Report output

RS transmitter error:
The receiver buffer of the
pH Meter contains an in-
complete character string
(line feed missing); this has
blocked pH Meter trans-
mission

Press <QUIT>

Send line feed

stirrer not
found

Various situations An attempt has been made

to control an 8xx series
stirrer which is not con-
nected

1. Press

<QUIT>

2. Switch off instrument
3. Check stirrer connection
or

Check measuring pa-

rameters

If necessary, change stir-

rer connection configura-
tion to a 7xx series stirrer
(see Section 5.6)

service is due Instrument has

been switched on

Set service date has been
reached

Contact service (see

Sec-

tion 5.5

)

change std
<ENTER>

Conc calibration

Instrument is waiting for a
new standard

1. Change standard
2. Press

<ENTER>

stop V reached Standard addi-

tion,
sample addition

Set stop volume has been
reached

1. Press

<QUIT>

2. Check method parame-

ters

system error
#X

System test after
switch on

pH/Ion Meter fault

1. Note error number
2. Contact Metrohm Service

----.- °C
or
check T-sensor

Various runs with
temperature
measurement

Temperature sensor is
faulty or has been un-
plugged during the run

Connect a properly function-
ing temperature sensor

Mode T

Temperature sensor is
faulty or not connected

Connect a properly function-
ing temperature sensor

transmission
error

PC keyboard
connecte

Communication problem
with PC keyboard

Check keyboard connection
with instrument switched off