8.6 pH electrode test

130

780/781 pH/Ion Meter, Manual

The following table lists the preset tolerances for the three types of elec-

trode:


excellent

electrode

good

electrode

passing

electrode

Electrode type: standard

Streaming potential dU [mV]

≤ 2.5

≤ 3.0

≤ 4.0

Total drift [mV]

≤ 2.0

≤ 2.5

≤ 3.0

Slope s [%]

96.5 ≤ s ≤ 101

96 ≤ s ≤ 102

95 ≤ s ≤ 103

Response time [s]

≤ 45

≤ 50

≤ 60

Offset potential U

off

[mV]

-15 ≤ U

off

≤ 15

Electrode type: gel => all (reference-) electrodes filled with Idrolyte

Streaming potential dU [mV]

≤ 3.0

≤ 3.5

≤ 4.5

Total drift [mV]

≤ 2.5

≤ 3.0

≤ 4.0

Slope s [%]

96.5 ≤ s ≤ 101

96 ≤ s ≤ 102

95 ≤ s ≤ 103

Response time [s]

≤ 60

≤ 75

≤ 90

Offset potential U

off

[mV]

-15 ≤ U

off

≤ 15

Electr. type: non-aqueous => all (reference-) electrodes filled with TEABr or LiCl in ethanol

Streaming potential dU [mV]

≤ 3

≤ 4.5

≤ 6

Total drift [mV]

≤ 5

≤ 7

≤ 9

Slope s [%]

88 ≤ s ≤ 120

80 ≤ s ≤ 130

70 ≤ s ≤ 140

Response time [s]

≤ 60

≤ 75

≤ 90

Offset potential U

off

[mV]

-10 ≤ U

off

≤ 70

The tested pH electrode is given that evaluation for which it fulfills all the

criteria. If one of the tolerances for a usable (passing) electrode is not

observed then the electrode test will be concluded with the message

'

electrode out of limits.

' in the report. If further test criteria are

also not fulfilled then the electrode test evaluation will be '

electrode

test not passed

'. In addition a more specific error message will be

shown on the display and will also be included in the report (see follow-

ing section).

TIP !

Instead of the stored tolerances for most Metrohm electrodes you can

set your own values (see Section 6.2.7). This may be helpful for spe-

cial applications if, for example, other quality requirements are placed

on the pH electrode or work is carried out with a reference electrode

which displaces the offset potential U

off

beyond the required –15 to

+15 mV (standard and gel electrodes).