6 standard methods – Metrohm 788 IC Filtration Sample Processor User Manual

Page 105

Advertising
background image

4.6 Standard methods

788 IC Filtration Sample Processor

97

4.6 Standard methods

The following pages contain listings of standard user methods included

with the instrument with explanations of the important commands. They

are required if the full control of the 788 IC Filtration Sample Processor

is not done by the PC directly.
Therefore, the standard methods are required for the combination of the

788 IC Filtration Sample Processor with the following Metrohm IC sys-

tems:

x 761 Compact IC or 790 Personal IC (see section. 2.6). These

IC systems work as "Master" and hold the time control of the IC

methods.

x Modular Metrohm IC Systems, connected to the PC by the 714

PC Board or 771 IC Compact Interface. Using the PC soft-

ware «

IC Net» or «IC Metrodata», a remote control via an addi-

tional serial connection is possible (see section 2.6).

For this configuration, two different versions exist: Either the 788

IC Filtration Sample Processor controls the time steps of the IC

method (788 as "Master") or the 732 IC Detector and the con-

nected PC, respectively, hold this time control (732 as "Master").

The following table briefly describes all standard methods of the 788 IC

Filtration Sample Processor. In the following sections, they are ex-

plained for their usage in combination with the IC systems mentioned

above. The specific IC methods have to be programmed at the 732 IC

Detector, via the PC program «IC Net» or via the PC programs for com-

pact IC systems, respectively.
Only with the methods 'SP ' and 'SP Seg' the 788 IC Filtration Sample

Processor works as "Master". All other methods are programmed for

usage of the 732 IC detector or the compact IC systems 761 and 790

as "Master".

Method Name

Function

PC

Processing a queue with the PC program

PC Seg Like

'PC',

additional insertion of an air bubble between the samples

SP

Processing a queue with the PC program with 788 as "Master"

SP Seg Like

'SP',

additional insertion of an air bubble between the samples

761

Processing a queue with 761 Compact IC or 790 Personal IC

761 Seg Like

'761',

additional insertion of an air bubble between the samples

An Cat

Simultaneous determination of anions and cations

AnCatSeg Like

'An Cat',

add. insertion of an air bubble between the samples

Preconc

Preconcentration of samples in an preconcentration column

Dialysis

Determination with automatic sample dialysis

Advertising