4 Operation

788 IC Filtration Sample Processor

108

4.7 Filtration

4.7.1 Selecting suitable sample types

Due to the small pore size of the filtration membrane, each filtration step

may contribute to possible clogging of the membrane.
In the following table some sample types are described, which have

been processed by the 788 IC Filtration Sample Processor. Filtration

was performed using the 6.2714.020 filtration membrane (0.15 µm) or a

second filtration membrane (0.2 µm). On a Metrohm IC System 7 stan-

dard anions were analyzed: F

-

, Cl

-

, NO

2

-

, Br

-

, NO

3

-

, HPO

4

2-

, SO

4

2-

.

Sample

Pore size of

filtration membrane

Number of samples

per membrane

orange juice with pulp

0.15

40

surface water

0.15

500

drinking water

0.15

1000

ground water

0.15

500

waste water 1

0.15

1000

waste water 2

0.15

130

waste water 3

0.15

40

waste water 4

0.15

80

NaCl solution (1 %)

0.2

5000

Schöninger decomp. sol.

0.2

100

acid soil extraction

0.2

1000

aqueous soil extraction

0.2

200


The number of samples, which can be filtrated on one membrane with-

out losses of analysis accuracy, is based on experience by Metrohm

and its customers. They are intended for evaluation of employment of

the ultra-filtration cell for automated sample pretreatment. These values

have to be determined for each new application individually.

4.7.2 Filtration membrane life

Analyzing standard solutions, a reducing recovery rate can be taken as

an indicator for possible clogging of the filtration membrane. These

standards are ideally prepared in the corresponding sample matrix.

Therefore, for the analysis of a large number of particle loaded sam-

ples, standard solutions should be included, e.g. after every fifth or

tenth sample.
Nevertheless, a prediction for the number of possible filtration steps on

one membrane cannot be made. Also the development of the recovery

rate may vary. The recovery rate could either be constant for a large

number of samples and drop abruptly or decrease continuously with

another type of sample matrix.