background image

4.2 Basic settings

788 IC Filtration Sample Processor

71

>rack definitions 2
type: M129-2

M129-2}

}

}

}

Type description of the rack

<SELECT> enables the choice of Metrohm-specific

and self-defined rack types (see section 4.5).

>rack definitions 2
work position 125 mm

0}125 mm

Working position of the needle

in mm from the upper stop.
Pressing <CLEAR> directly accepts the current lift po-

sition.

>rack definitions 2
rinse position 125 mm

0}125 mm

Rinsing position of the needle

in mm from the upper stop.
Pressing <CLEAR> directly accepts the current lift po-

sition.

>rack definitions 2
shift position 0 mm

0 mm

Shifting position of the needle

in mm from the upper stop.
Only the value

'0'

is permissible.

>rack definitions 2
special position 0 mm

0}125 mm

Special position of the needle

in mm from the upper stop.
Pressing <CLEAR> directly accepts the current lift po-

sition.

>rack definitions 2
>>special positions

Submenu

Special positions submenu

>>special positions
special beaker 1 128

0}number of positions

Position of special beaker 1

>>special positions
special beaker 2 129

0}number of positions

Position of special beaker 2

etc. up to special beaker 8

Up to 8 special beaker positions can be defined

(see section 4.5).