6 placing the sample rack, 7 adjusting the sample rack – Metrohm 788 IC Filtration Sample Processor User Manual

Page 22

Advertising
background image

2 Installation

788 IC Filtration Sample Processor

14

2222

4444

5555

6666

7777

8888

Fig. 6: Needle installation

If you are using the 6.1835.020 PEEK tubing, the sample tubes may

not be sealed with caps because they cannot be pierced by the PEEK

tubing and the needle may be damaged thereby.


2.3.6

Placing the sample rack

1

2

Place sample rack

Place sample rack 35

35

35

35 on the

turntable of the 788 IC Filtration

Sample Processor according

to Fig. 7.

Read magnetic code

Press <RESET> to move

the rack to the home position,

in which the magnetic code

can be read (details see

section 4.5).

Fig. 7: Sample rack placing


2.3.7

Adjusting the sample rack

If a new sample rack is placed on the 788 IC Filtration Sample

Processor for the first time, it must be adjusted on a rack position in the

middle row (example: position 37) as follows:

1 Place sample rack

Place empty sample rack on the 788 IC Filtration Sample

Processor and press <RESET> (see section 2.3.6).

2222 PEEK needle (6.1835.010) or

PEEK tubing (6.1835.020)

4444 Needle

holder

5555 PEEK compression fitting

(4.766.4320)

6666 PEEK compression fitting

(6.2744.010)

7777 Swing

head

8888 PTFE capillary (6.1803.070)