4 Operation

788 IC Filtration Sample Processor

70

>auxiliaries
beeper: ON

ON,OFF

Acoustic warning signal

ON

Acoustic warning signal in case of error

messages

OFF

No warning signal

>auxiliaries
device label

8 ASCII characters

Device

label

Freely selectable ASCII character string for characteriz-

ing the instrument (text entry, see section 4.1.6).

>auxiliaries
program 5.788.0010

read only

Number of the program version

Display only (no entry possibility).
Please specify this number in inquiries to Metrohm.

>auxiliaries
max. lift way 125 mm

0}125 mm

Maximum stroke path for needle

The following entry becomes effective after a RESET or

switching the IC Sample Processor off and on again.
This setting for max. lift way is

important for safe op-

eration. If the value for this entry is correct, damage to

the needle can be avoided because this prevents the

needle from being driven lower than the position indi-

cated. If you use the 6.2041.430 sample rack with

6.2743.050 sample tubes, the default value of 125 mm

should be retained.
Pressing <CLEAR> directly accepts the current lift po-

sition of the active tower.


configuration
>rack definitions

Definition of sample racks

>rack definitions
rack number 2

1}16

Number of the rack

The number of the rack in position is automatically dis-

played here, when its configuration is already stored

and when a RESET has been performed. If the configu-

ration of another rack must be changed, its rack num-

ber must be entered and confirmed with <ENTER>.

The rack number will be shown in the first menu line for

the subsequent entries. For more information concern-

ing sample racks, see section 4.5. The number

2

is as-

signed to the 6.2041.430 sample rack delivered.

>rack definitions 2
code 010001

6 bits

Identification code of the rack

The rack code must be unique and can only occur once

in the instrument (see section 4.5).