6 Appendix

788 IC Filtration Sample Processor

146

6.4.7 RS232

interface

This diagnosis step tests the functionality of the serial interface.

x Prepare instrument for diagnosis (see section 6.4.2).

x If necessary, press <Ð> several times until

diagnosis

>RS232 test

x <ENTER>

>RS232 test

RS232 test connector ?

x Insert the 3.496.8480 test plug into the RS232 interface without turn-

ing off the instrument.

x <ENTER>

The test runs automatically. If no error occurs, the following appears:

>RS232 test

RS232 test ok

x Remove test plug and press <ENTER>.

diagnosis

>external bus test




6.4.8

External Bus interface

This diagnosis step tests the functionality of parts of the internal EBUS

interface.

x Prepare instrument for diagnosis (see section 6.4.2).

x If necessary, press <Ð> several times until

diagnosis

>external bus test

x <ENTER>

If no error occurs, the following identification appears on the display:

>external bus test

address 0x86 type 3

x <ENTER>

>diagnosis

>beeper test