3 Operating tutorial

788 IC Filtration Sample Processor

52


<ENTER>

>sample sequence
8 WAIT 1200 s

x In this line a waiting time is defined that is used for the acquisition of

the chromatogram.

x The LEARN mode can also be used with the WAIT command.

<QUIT>

<Ð>

<ENTER>

>sample sequence
9 NOP

x An empty line with a

'NOP'

-entry (no operation) always forms the

end of a sequence.

x Exit the sample sequence with <QUIT> and go to the final se-

quence.



START

x After all sample tubes have been processed, the final sequence is

executed.

>final sequence
1 CTL:Rm: PUMP R/S 1

x In this line the pump drive at the 709 IC Pump is switched off. Press

<START>.


START

>final sequence
2 CTL:Rm: PUMP 752 OFF

x In this line the pump drive at the 752 Pump Unit is switched off.

Press <START>.


QUIT

>final sequence
3 NOP

x Now you have reached the end of the final sequence and have

completed the entire run of a sample series.

x By pressing <QUIT> twice the 788 IC Filtration Sample Processor

returns to the normal state.



PARAM

7

SAMPLE

START

x Now prepare some sample tubes and place them on the sample

rack. Prepare the IC system with 732 IC Detector, 733 IC Separation

Center, 709 IC Pump and 752 Pump Unit for the recording of chro-

matograms.

x Enter the number of samples to be processed (<PARAM>) and

define the position of the first sample (SAMPLE = 1).

x Now you can start your first sample series with <START>.