Metrohm 788 IC Filtration Sample Processor User Manual

Page 59

Advertising
background image

3.5 "Tracing"

788 IC Filtration Sample Processor

51

HOLD

LEARN

<ENTER>

x Press <LEARN> to access the LEARN mode. The blinking LEARN-

LED indicates that the 788 IC Filtration Sample Processor is ready to

execute the command.

x Now move the lift into the desired position with the <Ð> and <Ï>

keys. You will notice that the current lift position is always indicated

"live". During execution of the command the LEARN-LED is lit con-

tinuously. Accept the lift position that has been set by pressing

<ENTER> and thereby exit the LEARN mode. The LEARN LED

goes off again.

<START>

>sample sequence
4 CTL:Rm: FILL A 1


x In this line the injection valve at the 733 IC Separation Center is

switched to the "Fill" position.

<START>

HOLD

LEARN

HOLD

LEARN



<ENTER>

>sample sequence
5 PUMP 1.1 : 240 s

x In this line the peristaltic pump at the 788 IC Filtration Sample Proc-

essor is started in order to convey the sample from the sample tube

to the sample loop at the 733 IC Separation Center.

x Here, you can use the LEARN mode to optimize the pump time too.
x In this case, as with the other "teachable" commands (the LIFT

command is an exception), pressing the <LEARN> key causes

immediate execution of the corresponding command. The elapsed

time is also displayed here "live". By pressing the <LEARN> key

again the command can be interrupted.

x The blinking LED indicates that the 788 IC Filtration Sample Proces-

sor is still in the LEARN mode. If you now switch the pump back on

with the <LEARN> key, you will see that the "live" value (pump time)

is now added to the existing value.

x Now set the pump time in this way. Accept the total time with

<ENTER> and exit the LEARN mode in this way.

<START>

>sample sequence
6 CTL:Rm: ZERO 1

x In this line the autozero function at the 732 IC Detector is triggered.

<START>

>sample sequence
7 CTL:Rm: INJECT A 1

x In this line the injection valve at the 733 IC Separation Center is

switched to the "Inject" position.