4 manual operation, 2 moving the lift – Metrohm 788 IC Filtration Sample Processor User Manual

Page 96

Advertising
background image

4 Operation

788 IC Filtration Sample Processor

88

4.4 Manual

operation

4.4.1

Turning the sample rack / Positioning the samples

Using the <Í> and <Î> keys the sample rack can be

turned one position to the left (in the counterclockwise direction)

or right (clockwise).

8

MOVE

manual operation
MOVE 1 : sample

sample,spec.1}8,

1}999

Position vessel /

Turn rack

With the MOVE command, a particular sample tube or other

vessel can be placed under the needle. With <SELECT> the

numerical rack position as well as the predefined current sam-

ple (SAMPLE command) or the special beakers 1-8 can be cho-

sen.

For security reasons turning the sample rack is only possible

when the lift is in or above the shift position.


4.4.2

Moving the lift

The keys <Ï> and <Ð> allow upward and downward move-

ment of the lift. The lowest possible lift position is defined by the

configuration parameter

'max. lift way'

.

HOME

END

The <HOME> key runs the lift to the rest position (0 mm), i.e. to

the upper limit.
<END> runs the lift to the predefined work position (see sec-

tion 4.3.2).

9

LIFT

manual operation
LIFT: 1 : rest mm

work,rinse,shift,

special,rest,

0}125 mm

Position the lift

With the LIFT command, the lift can be run to a given position.

In addition to selecting an exact position in mm (0 - 125 mm),

the <SELECT> key can select a predefined position (work

position, rinse position, shift position, special position, rest

position = 0 mm).