5 error messages – Metrohm 788 IC Filtration Sample Processor User Manual

Page 127

Advertising
background image

5.2 RS232 interface

788 IC Filtration Sample Processor

119

5.2.5 Error

messages

Error messages ‘

EXXX

’ are appended to the status message and sepa-

rated from this by a ‘

;

’.

Error

Meaning

Exit/Corrective action

E1

Incorrect program checksum

Inform Metrohm service.

E2

RAM read/write error

Inform Metrohm service.

E3

RAM lost data

<QUIT>. If instrument adjustment is neces-
sary, inform Metrohm service.

E4

Timer interrupt for multi-tasking missing

Inform Metrohm service.

E5

RS232 module test error

Inform Metrohm service.

E6

RS232 read/write error

Inform Metrohm service.

E7

Display read/write error

Inform Metrohm service.

E18

Low battery

Inform Metrohm service.

E19

RAM test error

Inform Metrohm service.

E28

Wrong object call-up.

Correct path.

E29

Wrong value.

Enter correct value or new path.

E30

Wrong trigger.

Enter correct trigger or new path.

E36

RS232 receive error; parity.

<QUIT>, set same parity for both devices.

E37

RS232 receive error; stop bit.

<QUIT>, set same stop bit for both devices.

E38

RS232 receive error; overflow (at least 1 character
could not be read).

<QUIT>, set same baud rate for both de-
vices, restart sender.

E39

RS232 receive error; overflow of internal receive
buffer (>82 characters).

<QUIT>.

E40

RS232 send error; DSR=OFF. Handshake not
answered for longer than 1 s.

<QUIT>, check receiver
(switched on and ready?).

E41

RS232 send error; DCD=ON.
Handshake not answered for longer than 1 s.

<QUIT>, check receiver
(switched on and ready?).

E42

RS232 send error; CTS=OFF. Handshake not
answered for more than 1 s.

<QUIT>, check receiver
(switched on and ready?).

E43

RS232 send error; transmission of the IC Filtration
Sample Processor was interrupted with XOFF for
at least 3 s.

<QUIT> or send XON.

E44

RS232 send error; the RS parameters are no
longer the same for both devices.

<QUIT>, reset RS parameters for both de-
vices.

E45

RS232 send error; the receive buffer of the 788 IC
Filtration Sample Processor contains an incom-
plete string (L

F

FF

F

missing), the transmission is thus

blocked.

<QUIT> or send L

F

FF

F

.

E50}E59 Error in IO test.

<QUIT>, inform Metrohm service.

E60}E82 Error in RS232 test.

<QUIT>, check connection between the RS
interfaces.

E201

Functional error.

Inform Metrohm service.

Advertising