4 Operation

788 IC Filtration Sample Processor

72

configuration
>RS232 settings

Settings for RS232 interface

For further details on the RS232 interface, see section 5.2.

>RS232 settings
baud rate: 9600

9600,4800,2400,

1200,600,300

Data transmission rate (baud rate)

Data transmission rate in bit/s

>RS232 settings
data bit: 8

7,8

Data

bits

>RS232 settings
stop bit: 1

1,2

Stop

bits

>RS232 settings
parity: none

none,odd,even

Parity

none

Parity is not checked.

odd

Odd

parity.

even

Even

parity.

>RS232 settings
handshake: HWs

HWs,HWf,

SWchar,SWline,none

Handshake

HWs

Reduced hardware handshake.

HWf

Full hardware handshake.

SWchar

Software handshake with character stop.

SWline

Software handshake with line stop.

none

No

handshake.

For detailed information on the handshake, see section

5.2.8.

>RS232 settings
character set: IBM

IBM,HP,Epson,
Seiko,Citizen

Character set for printer or PC

The settings for the printers recommended by Metrohm

are listed in section 2.7.2. For printers not listed, the set-

ting

'Epson'

is recommended. In any case the printer

handbook should be consulted. For data transfer with

personal computers,

'IBM'

must be chosen.

>RS232 settings
RS control: ON

ON,OFF

Control via RS232 interface

ON

Data receipt via RS232 interface switched

on.

OFF

Data receipt via RS232 interface switched off

(no external control via RS232 possible).