6 remote interface – Metrohm 788 IC Filtration Sample Processor User Manual
Page 153
6.4 Diagnosis
788 IC Filtration Sample Processor
145
Key code table
Code
Key
Code
Key
0
16
<7 / SAMPLE>
1
<HOLD / LEARN>
17
<4 / PUMP>
2
<STOP>
18
<1 / SCAN>
3 <START> 19 <0>
4 <CONFIG> 20 <END>
5 <PARAM> 21 <Î>
6
<USER METHOD>
22
<CLEAR / RESET>
7 23
<ENTER>
8
<9 / LIFT>
24
<Ï>
9 <6> 25 <Ð>
10
<3 / WAIT>
26
<SELECT / TOWER>
11
<* / ENDSEQ>
27
<QUIT>
12
<8 / MOVE>
28
<HOME>
13 <5> 29 <Í>
14
<2 / CTRL>
30
<INSERT >
15
<. / PRINT>
31
<DELETE>
6.4.6 Remote
interface
This diagnosis step tests the functionality of all the output (14) and input
lines (8).
x Prepare instrument for diagnosis (see section 6.4.2).
x If necessary, press <Ð> several times until
diagnosis
>remote test
x <ENTER>
>remote test
remote test connector ?
x Plug 3.496.8550 test plug into the remote interface without turning
the instrument off.
x <ENTER>
The test runs automatically. If no error occurs, the following appears:
>remote test
remote test ok
x Remove test plug and press <ENTER>.
diagnosis
>RS232 test