6.4 Diagnosis

788 IC Filtration Sample Processor

145

Key code table

Code

Key

Code

Key

0

16

<7 / SAMPLE>

1

<HOLD / LEARN>

17

<4 / PUMP>

2

<STOP>

18

<1 / SCAN>

3 <START> 19 <0>
4 <CONFIG> 20 <END>
5 <PARAM> 21 <Î>
6

<USER METHOD>

22

<CLEAR / RESET>

7 23

<ENTER>

8

<9 / LIFT>

24

<Ï>

9 <6> 25 <Ð>

10

<3 / WAIT>

26

<SELECT / TOWER>

11

<* / ENDSEQ>

27

<QUIT>

12

<8 / MOVE>

28

<HOME>

13 <5> 29 <Í>
14

<2 / CTRL>

30

<INSERT >

15

<. / PRINT>

31

<DELETE>

6.4.6 Remote

interface

This diagnosis step tests the functionality of all the output (14) and input

lines (8).

x Prepare instrument for diagnosis (see section 6.4.2).

x If necessary, press <Ð> several times until

diagnosis

>remote test

x <ENTER>

>remote test

remote test connector ?

x Plug 3.496.8550 test plug into the remote interface without turning

the instrument off.

x <ENTER>

The test runs automatically. If no error occurs, the following appears:

>remote test

remote test ok

x Remove test plug and press <ENTER>.

diagnosis

>RS232 test