9 beeper, 10 rack code recognition – Metrohm 788 IC Filtration Sample Processor User Manual

Page 155

Advertising
background image

6.4 Diagnosis

788 IC Filtration Sample Processor

147

6.4.9 Beeper

x Prepare instrument for diagnosis (see section 6.4.2).

x If necessary, press <Ð> several times until

diagnosis

>beeper test

x <ENTER>

The beeper is turned on and off in an endless loop.

>beeper test

x The test can be exited by pressing the <QUIT> or <STOP> keys.

diagnosis

>rack code test


6.4.10 Rack code recognition

This diagnosis step checks the functionality of the sensors with regard

to their ability to automatically recognize the attached rack.

x Note the code (magnets) on the underside of all the racks being

used. The following diagram shows the magnet fixture on the under-

side of a rack.

Magnets inserted

In this example
The code is:

000110

Magnet fixture as viewed from underneath

x Pick up the rack and lay it on its side.

x Prepare the instrument for diagnosis (see section 6.4.2).

x If necessary press <Ð> several times until

diagnosis

>power on reset

x <ENTER>

The sample changer runs through its start routine (initializing lift and rack positions). The initiali-
zation is important because it is vital for the following diagnosis step “rack code test“ that the
rack is in the basic position.

x Prepare instrument for diagnosis (see section 6.4.2).

x If necessary press <Ð> several times until

diagnosis

>rack code test