2.4 Installation for full control with «IC Net»

788 IC Filtration Sample Processor

27

2.4.5 MIC

2

The 788 IC Filtration Sample Processor is connected to an anion IC

system with chemical suppression consisting of 732 IC Detector, 733 IC

Separation Center, 709 IC Pump, 752 IC Pump Unit, and 762 IC

Interface as shown in Fig. 14 using the 6.2141.110 remote connection

cable and the 6.2134.080 RS232 cable.

762

733

709

732

PC

6.2134.100

6.2125.090

788

6.2134.080

6.2141.110

752

6.2134.090

6.2128.130

6.2115.070

6.2128.180

Fig. 14: Interconnection with anion IC system with chemical

suppression (MIC 2)

Settings in the «IC Net 2.1» program


Creation of a time program for the 788 IC Filtration Sample Processor:

001 Ctrl

INIT

732

002 Move sample
003 Lift

work

004 Ctrl

ZERO

1

005 Ctrl

FILL A 1

006 Pump 240 s
007 Ctrl

INJECT A 1

 Initialize remote interface at 732
 Move needle to sample
 Place lift with needle to working position
 Trigger autozero at 732
 Switch injection valve A at 733 to "Fill"
 Fill sample loop with sample during 240 s
 Switch injection valve A at 733 to "Inject"

Modular anion system with chemical suppression