2.7 Connection of devices to the RS232 interface

788 IC Filtration Sample Processor

41

2.7

Connection of devices to the RS232
interface

2.7.1

General information on RS232 interface

Many different instruments may be connected via the serial RS232

interface 40

40

40

40. In addition to all Metrohm instruments that support the

Metrohm remote control language (see section 5.2) any printer with

serial interface (or parallel interface and parallel/serial converter) or a

personal computer (PC) may be connected. Any other measuring

instrument may be controlled via RS232 interface, as long as it supports

serial data transmission.

732

788

Fig. 23: Connection possibilities for the RS232 interface

In order to guarantee safe data transmission, it is important to set the

same RS232 interface parameters correctly for both instruments

connected (see section 2.5.2).

Control commands (examples)

CTL:RS &M;$G

starts a Metrohm instrument

CTL:RS &M;$S

stops a Metrohm instrument

PRINT: config

prints a configuration report to

a printer or PC

Scanning input data (example):

SCN:RS : *R"

waiting for readiness of a Metrohm

instrument

The settings and cables for the connection of a printer to the 788 IC

Filtration Sample Processor are described in section 2.7.2.

6.2125.060 cable

Printer cables, see section 2.5.2