6.3 Maintenance and servicing

788 IC Filtration Sample Processor

139

6.3.4

Replacing the filtration membrane

A good condition of the filtration membrane is a prerequisite for a con-

stant quality of the analysis results. Therefore, the membrane has to be

replaced regularly.
Please consider section 4.7 for evaluating the membrane type and time

of exchange.
Replace the filtration membrane as follows, see also Fig. 9 in section

2.3.8.

1 Prepare ultra-filtration cell

x Remove all tubings at the ultra-filtration cell 34

34

34

34 by opening all

PVDF compression fittings 22

22

22

22.

x Using the 6.2621.070 Allen key, completely loosen the 5

screws 46

46

46

46, separate top part 43

43

43

43 from bottom part 45

45

45

45 and re-

move sealing ring 44

44

44

44.

x Thoroughly rinse the sealing ring, bottom part and top part of

the dialysis cell with ultrapure water.

Use only ultrapure water or ethanol to clean the dialysis cell; or-

ganic solvents (e.g. acetone) will cause damage to the plexiglas cell!


2 Prepare filtration membrane

x Use the 6.2831.010 tweezers to extract a new filtration mem-

brane 50

50

50

50 (6.2714.020) from its packaging and immerse in a

Petri dish containing ultra pure water for approx. 2 min until

the membrane is completely saturated with water.

Because the filtration membrane it is stretched out thereby, it has to

be watered before inserting into the ultra-filtration cell. Otherwise, the

membrane could be crumbled in the cell, which leads to blocking.

3 Insert filtration membrane

x Lay top part 43

43

43

43 on a paper wipe with its inside facing up-

wards.

x Insert sealing ring 44

44

44

44 in the appropriate recess in the top part.

x Using the tweezers, place the wet dialysis membrane inside

the sealing ring on the top part of the filtration cell.