6 commands for sequences – Metrohm 788 IC Filtration Sample Processor User Manual

Page 88

Advertising
background image

4 Operation

788 IC Filtration Sample Processor

80

4.3.6

Commands for sequences

The following commands are programmable within a sequence. Most of

them are also available in manual operation but may vary partially in

their operation or exhibit a limited parameter selection (see section 4.4).

SAMPLE

7

SAMPLE

>start sequence
1 SAMPLE: = 1

=,+,–; 1}999

Define current sample

position

The SAMPLE command defines the rack position for the current sample

(

SAMPLE = X

). This is stored as a index variable. It may be modified for

example, in a sample sequence (

SAMPLE + X

or

SAMPLE – X

), in order to

control the course of a sample series during processing.
The SAMPLE command does not have to be used for simple applica-

tions. As a general rule the first sample of a series is assumed to be in

rack position 1 unless specified otherwise. Therefore it is recommended

not to place the special beakers in the first rack positions; place them in

the highest positions instead.
Before starting a sample series, the position of the first sample can be

defined with the <SAMPLE> key in manual operation, as long as this

has not been defined in the method itself.
If a certain sample tube order is needed for each application of a

method, the position of the first sample can be defined in the start se-

quence with

'SAMPLE = X'

and this setting can be saved with the corre-

sponding method.
If the SAMPLE command is not executed during a sample sequence,

the SAMPLE variable will be increased by 1 after every processing of

the sample sequence.

MOVE

8

MOVE

>sample sequence
2 MOVE 1 : sample

sample,spec.1}8,1}999

Position vessel /

Turn rack

The MOVE command can place the current sample tube or a special

beaker below the needle position by rotating the rack. An absolute rack

position can also be specified.
During method processing a MOVE command automatically sets the lift

to the shift position. In the parameter menu under

'>changer settings'

the turning speed can be defined specifically for each method.

If there is no sample tube in the rack position chosen, this is not

recognized by the IC Sample Processor, and air is aspirated instead of

sample solution. Therefore always make sure that sample tubes are

placed at all rack positions defined in the processing sequence.