6 Appendix

788 IC Filtration Sample Processor

144

diagnosis

>display contrast test

x <ENTER>

After pressing the <ENTER> key the following display appears whereby the contrast of the dis-
play continuously varies between bright and dark.

>display contrast test

788 IC Filtr. Processor

x The test can be exited with the <QUIT> or <STOP> key.

diagnosis

>key test







6.4.5 Keyboard

This diagnosis step allows all the keys on the keyboard to be tested for

their functionality.

x Prepare instrument for diagnosis (see section 6.4.2).

x If necessary, press <Ð> several times until

diagnosis

>key test

x <ENTER>

>key test

x Activate all the keys one after the other and check their reaction on

the display.

The corresponding matrix code appears in the display and a description of the main function of
the key pressed (for example, the following display should appear when the <CONFIG> key is
pressed).

>key test

code 4 CONFIG

x The test is exited by pressing the <STOP> key twice.

diagnosis

>remote test