5.2 RS232 interface

788 IC Filtration Sample Processor

127

5.2.7

Data transmission protocol

The RS232 interface of the 788 IC Filtration Sample Processor is con-

figured as DTE (Data Terminal Equipment) with the following technical

specifications:

x Standard

Data interface in accordance with EIA standard

RS 232C (DIN 66020, page 1), transmission pa-

rameters can be set under

'configura-

tion/RS232 settings'

, see section 4.2.1.

x Control characters C

R

DEC 13

HEX 0D

L

F

DEC 10

HEX 0A

XON

DEC 17

HEX 11

XOFF

DEC 19

HEX 13

x Max. line length

80 characters + C

R

L

F

x Mode

Full duplex (simultaneous transmission and re-

ceiving)
Restriction:

If data are first received by the interface, trans-

mission is not started until the receipt is at an

end.

x Cable

For interconnection of the 788 IC Filtration

Sample Processor with third-party devices, only

a shielded data cable may be used (e.g.

METROHM D.104.0201). The cable shield must

be perfectly earthed at both devices (pay atten-

tion to current loops: always use star earthing).

Use only connectors with sufficient shielding

(e.g. Metrohm K.210.0001 with K.210.9004).

x Cable length

max. ca. 15 m

Start

7 or 8 data bits

parity bit

1 or 2 stop bits

Advertising