4.6.6 Method "Dialysis"

Application

IC determinations with automatic sample dialysis. This method is not

planned for usage together with the ultra-filtration cell, which should be

bridged. The present method contains a simultaneous separation of one

sample and the dialysis of the following sample within the sample se-

quence. The start sequence just starts the dialysis of the first sample

while the final sequence only starts the determination of the final sample.

Interconnection

See section 2.6.5 as well as 754 Instructions for Use

Program of 788 IC Filtration Sample Processor

788 IC Filtr. Proc. 5.788.0010
parameters
method Dialysis
number of sample: rack
>start sequence
1 CTL:Rm: INIT
2 CTL:Rm: PUMP 752 ON
3 MOVE 1 : 1
4 LIFT: 1 : work mm
5 CTL:Rm: PUMP 754 ON
6 WAIT 120 s
7 CTL:Rm: FILL B/STEP 1
8 WAIT 600 s
>sample sequence
1 SCN:Rm : Pump1 ?
2 CTL:Rm: ZERO 1
3 CTL:Rm: STEP MSM 753
4 CTL:Rm: FILL A 1
5 CTL:Rm: INJECT B 1
6 WAIT 30 s
7 CTL:Rm: INJECT A 1
8 CTL:Rm: PUMP 754 OFF
9 SCN:Rm : Wait1
10 SAMPLE: + 1
11 MOVE 1 : sample
12 LIFT: 1 : work mm
13 CTL:Rm: PUMP 754 ON
14 WAIT 120 s
15 CTL:Rm: FILL B/STEP 1
16 WAIT 600 s
>final sequence
1 CTL:Rm: ZERO 1
2 CTL:Rm: STEP MSM 753
3 CTL:Rm: FILL A 1
4 CTL:Rm: INJECT B 1
5 WAIT 30 s
6 CTL:Rm: INJECT A 1
7 CTL:Rm: PUMP 754 OFF
8 WAIT 1200 s
9 CTL:Rm: PUMP 752 OFF
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:

------------

 Report header with serial number and program ver-

sion

 Method name
 Number of samples to be processed (entire sample

rack)

 Initialize remote interface
 Switch 752 Pump Unit on
 Move needle to sample position
 Place lift with needle to working position
 Switch 754 Pump Unit on
 Rinse dialysis cell
 Stop pumping acceptor solution
 Dialysis time


 Waiting until 709 IC Pump runs
 Trigger autozero at 732 IC Detector
 Switch 753 suppressor module to next position
 Switch injection valve A at 733 to "Fill"
 Transfer of acceptor solution to sample loop
 Wait 30 s
 Switch injection valve A at 733 to "Inject"
 Switch 754 Pump Unit off
 Waiting until 732 IC Det. sends signal on rem. line 3
 Raise sample position by 1
 Move needle to sample position
 Place lift with needle to working position
 Switch 754 Pump Unit on
 Rinse dialysis cell
 Stop pumping acceptor solution
 Dialysis time


 Trigger autozero at 732 IC Detector
 Switch 753 suppressor module to next position
 Switch injection valve A at 733 to "Fill"
 Transfer of acceptor solution to sample loop
 Wait 30 s
 Switch injection valve A at 733 to "Inject"
 Switch 754 Pump Unit off
 Wait for final suppressor regeneration
 Switch 752 Pump Unit off