2 Installation

788 IC Filtration Sample Processor

36

Settings in the «790 Personal IC» program

A time program must be drawn up for the selected system which first

produces an impulse at remote lead 3 to start sample changing on the

788 IC Filtration Sample Processor and then switches the injection

valve to the "Fill" position. A further impulse is then produced to start

filling the sample loop within 240 s. At the end the injection valve is

switched to the "Inject" position, which also starts data acquisition.

Run conditions

In order that the joint operation of 790 – 788 functions properly the

program must be started first on the 788 IC Filtration Sample

Processor, then the "Sample Queue" on 790 Personal IC.