3.3 Rack configuration

788 IC Filtration Sample Processor

47

RESET

CLEAR

x At the end of this basic configuration the 788 IC Filtration Sample

Processor must be turned off and on again or re-initialized by press-

ing <RESET> to make the latest settings effective.

x All data entered up to this point however, are retained. The same is

true for any methods that may have been saved.

3.3 Rack

configuration

x Using the keys <Ï> and <Ð> you can run the lift to the desired

work position for the needle.

CONFIG

<Ð>

<ENTER>

x Now open the configuration menu with <CONFIG> and move the

cursor key <Ð>, until you reach the submenu

'>rack defini-

tions'

. Press <ENTER> to open this submenu where you can de-

fine the rack configuration.

>rack definitions
rack number 2

<ENTER>

x The rack number of the engaged rack will be displayed as soon as

the sample rack has been correctly identified. By confirming with

<ENTER> you access the rack data. (By entering another rack

number you can also edit the data of a sample rack that is not en-

gaged.)

<Ð>



RESET

CLEAR




x You can skip the first entries (code and rack type) with the cursor

key <Ð>. Now you can enter the work position of the needle.

>rack definitions 2
work position 125 mm

x Because you have already positioned the needle at the desired

height, you can accept the current lift position directly by pressing

<CLEAR>. Of course the work position can be entered manually or

the value that has automatically been accepted can be modified

later. Lift positions are given in millimeters (0…125 mm), measured

from the uppermost limit (rest position) of the lift. Consider during

the input of the working position that with sample tubes sealed with

non-perforated PE caps the work position must be set to 125 mm,

since otherwise a vacuum can develop in the sample tube and the

sample will not be aspirated correctly.