3 replacing the pump tubing – Metrohm 788 IC Filtration Sample Processor User Manual

Page 146

Advertising
background image

6 Appendix

788 IC Filtration Sample Processor

138

6.3.3

Replacing the pump tubing

Pump tubings are consumable material with a limited lifetime and

should be replaced at regular intervals (approx. every 2 weeks under

continuous use).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be correctly

set as described in section 2.3.9. If the pump is to remain switched off

for a lengthy period of time the tubing cartridges should be raised

completely by loosening snap-action lever 16

16

16

16 on the right-hand side

(the pre-set contact pressure remains unchanged).
To replace a pump tubing proceed as follows:

1 Remove old pump tubing

x Press contact pressure lever 18

18

18

18 on the tubing cartridge down

as far as it will go.

x Release tubing cartridge 14

14

14

14 from holding clamp 17

17

17

17 by press-

ing down snap-action lever 16

16

16

16 and remove from mounting pin

13

13

13

13 at the 788 IC Filtration Sample Processor (see Fig. 1).

x Remove old pump tubing.

2 Insert new pump tubing

x Insert the new pump tubing 10

10

10

10 (6.1826.070) or 11

11

11

11

(6.1826.030) in the tubing cartridge as shown in Fig. 10. The

stopper 43

43

43

43 must click into the corresponding holder on the

left-hand side of the tubing cartridge.

x Place the tubing cartridge on mounting pin 13

13

13

13 and press

down on the right-hand side until snap-action lever 16

16

16

16 clicks

into position on holding clamp 17

17

17

17. Take care that no kinks are

formed in the pump tubing.

3 Set contact pressure

x Press contact pressure lever 18

18

18

18 upwards until the solution

just starts to be drawn in. Then press contact pressure lever

upwards until it clicks once more to obtain optimal contact

pressure.


The 6.1826.0X0 pump tubing is made of PVC and must not be used

for rinsing with solutions which contain acetone. In such cases, rinse

with different pump tubing or a different pump.